en - Areas of application
38
Test point for measuring water pressure
Looking at the water connection at the back, the 4th connection on
the water pipe can be used as the test point for measuring water
pressure for performance checks and for validation according to EN
ISO 15883.
To measure water pressure replace the existing blind stopper with a
Luer Lock adapter, e.g. E 447.