Electrical lift units
g
2
1
1
2
2
12
Lift unit for
worktop with
safety rail
cannot be height-
adjusted.
1
1
Fault in safety system.
The worktop cannot run up or down if the
safety rail is defective.
Check that all wires and plugs are intact
and correctly connected according to the
installation guide.
Assembly error on
Use measurement device
safety rail system.
to detect current. Measure current on
the wire from the safety rail and its plug
(a close function).
- if there is current, the safety device
system is working. The lift unit should now
run up/down. REMEMBER to reset.
- if no current can be detected:
Remove the plastic corners on the safety
rails and slacken the screws at both ends.
The rails should now move freely. Note that
the rails' freedom to move can be hindered
by dirt or grease.
Check for current once more:
If there is current, the safety device system
is working. If no current can be detected,
see under "Faulty switches".
Faulty switches.
Check whether one or more of the safety
system contacts is sticking, and if so,
attempt repair . If the problem cannot be
fi xed, replace the contact(s). The safety
system will only work when current can be
detected. REMEMBER to reset.