Function And Use Of Buttons - Hydas 4507.1.00 Mode D'emploi

Appareil d'électro stimulation - «easy free» tens
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
4. Finishing the operation
You should hold the plug when pulling it out. Please do not pull the wire.
• Please use clear water to wash or use wet cloth (instead of facial
tissue) to gently wipe up the electrodes when cleaning them.
• Wires stay away from babies and children, prevent the winding neck
causing suffocation and death.

6.1 | Function and use of buttons

[ON]-button:
• Press this button to turn on the unit and increase the output
intensity.
• Each time you press the [ON]-button the intensity of
stimulation will increase by one level. (0 - 8, 8 levels in total)
[OFF]-button:
• Press this button to turn off the unit and decrease the output
intensity.
• Each time you press the [OFF]-button the intensity of
stimulation will decrease by one level (8 - 0, 8 levels in total)
• The output intensity level will be displayed by the number on the LCD
display: «1» is the lowest, «8» is highest. The output strength will gradually
increase according to the selection of intensity level.
• If the output intensity is at «0», it means that the device is standby status
and will automatically shut off in 20 seconds.
• When the output intensity is at 1- 8, the time setting function starts to work.
The black spot on the LCD display will flash once each second.
• The adhesive pads should be attached around the area of pain before
setting the stimulation mode and output intensity.
[MODE/STOP]-button:
• To adjust the output function to a suitable level, press the
[MODE/STOP]-button
• There are 8 different selections of stimulation settings (form
A - H). For detailed descriptions please refer to the following
function chart.
The device has a memory function. Once a particular stimulation setting is chosen
it will remain on that setting even when the device is switched off and switched back
on. For example, if the device was set to «B» mode before being switched off, when
the device is restarted, it will remain in «B» mode until another mode is chosen.