1. Device Cover
CardiAid is switched on by opening the cover and switched off by closing the cover.
Closing the cover will switch off the device only after the electrodes are disconnected.
2. Socket for Electrode Plug
Electrodes are connected to CardiAid through this socket. Electrodes supplied with the device are
already connected to this socket.
3. Loudspeaker
Audio warnings of the CardiAid are heard through this loudspeaker.
4. Emergency Kit
Emergency kit includes scissors, shaver, respiration mask, gloves, and alcohol pad.
Emergency kit should be replaced after use.
5. Defibrillation Electrodes
Electroshock is delivered to the patient through these electrodes. Electrodes should be replaced
after each use.
6. Protection Bag
Protection bag is used to store, carry, and protect the device.
7. Indicator Window
The status of the device can be observed through this window on the protection bag.
8. Indicator: "Remove clothing from chest and stick on electrodes."
When this indicator lights, you should stick the electrodes on patient's bare chest.
9. Indicator: "Do not touch the patient from now on."
Patient should not be touched when this indicator is flashing. For example: During heart rhythm
analysis and shock application.
10. Indicator: "Patient may be touched."
While this indicator is lighting, patient may be touched. For example: During
cardiopulmonary resuscitation.
11. Indicator: Ready for Shock (only for CT0207RS)
This indicator flashes when CardiAid is ready to deliver shock.
12. Shock Button (only for CT0207RS)
Shock button starts flashing after shock is prepared. This button is pressed to deliver electroshock.
13. Info-button
When pressed, an audible notification indicating the duration of use and number of shocks applied
is heard.
14. Repair Symbol
The device should not be used, if the repair symbol is flashing or lighting. In this case, it should be
repaired by Cardia International or an authorized service provider of Cardia International.
15. Battery Symbol
The device is not ready for use, if the battery symbol is flashing or lighting. In this case, contact
Cardia International or an authorized service provider of Cardia International immediately.
16. OK Symbol
CardiAid is ready for use, if the OK symbol is flashing while the device is switched off.
Overview of CardiAid CT0207RS & CT0207RF
5