Section 3 - Operator Control Panel
32°C
3.0 Operator Control Panel
The operator controls are identified in figure 4.
3.1 POWER Switch
The POWER switch controls power to the Power
Unit. The POWER INDICATOR light in the switch
will be lit when power is on.
3.2 TEMPERATURE SETTING
Push Buttons
When the Power Unit is turned on, the system
defaults to the 32°C temperature setting.
Temperature settings are selected by pressing the
appropriate push button-- fan only, 32°C, 38°C,
43°C, or 46°C.
If electrical power is momentarily interrupted,
the existing temperature setting will remain. If
electrical power remains off, the TC3003 will
default to 32°C when power is restored.
38°C
43°C
Figure 4 - Operator Control Panel
Failure to follow the following
precautions could result in death or
serious injury:
• Do not use the 46°C setting in the
• Do not use the 46°C setting for
3.3 REMOVE FROM USE Indicator
The REMOVE FROM USE indicator will light:
• if a patient limit thermostat or heater
overtemp thermostat has tripped, or
• if the HEPA filter has been installed
incorrectly or is missing.
The heating element and blower will automatically
turn off when either of these conditions occur.
If the REMOVE FROM USE indicator
lights, turn the Power Unit off
immediately. Discontinue use and refer
equipment for servicing.
46°C
WARNING
operating room.
patients with any of the following
conditions:
1. Low cardiac output
2. Poor peripheral perfusion or
peripheral vascular disease
3. Total immobilization
4. Non-sensate patients
CAUTION
5