7 Trouble-shooting
The following table shows explanations of the error messages and tips on how to
avoid incorrect measurements.
The most commonly encountered problems are caused by:
•
The test strips or cells not being inserted correctly.
•
Cells being dirty or strongly scratched.
•
The reaction time of the strip not being right.
•
Incorrect use of the tests (e. g. strips / tests not properly stored, pH range
not cor
rectly adjusted).
Problem
Cause
• No display
• Batteries not
en
• Batteries are drained
• Display LO
• Concentration is
• Display HI
• Concentration is hig-
• Display - - -
• No calibration pro-
|
28
Solution
• Insert the batteries
inserted properly
correctly
• Use new batteries
• State the result is
lower than the meas-
smaller than the
urement range of the
lowest measurement
test
value
• Dilute the sample; do
her than the measu-
not forget to consider
rement range of the
the factor for dilution
test
when reporting the
result!
• Calibrate instrument
grammed
Problem
Cause
• Display OPT
• Optics or strip adap-
ter is dirty
• Display ERR
• Optical error
• Strip adapter is
assembled incorrectly
• Display E-1
• Surrounding is too
bright
• Strip adapter is
assembled incorrectly
• Display E-2
• Result is not plausible
Solution
• Clean optics and
adapter and switch
the RQflex® plus 10
on and then off
again; recalibrate, see
page 26
• Clean optics and
adapter; recalibrate
instrument, see page
26; if necessary, call
customer service
• Repeat measurement
in a darker place;
en
if necessary, close
the lid of the instru-
ment before switching
on
• Repeat measurement
|
29