Recommended separation distance between Portable and mobile RF
communications equipment and HemoCue
The HemoCue systems are intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
The customer or the user of HemoCue systems can help prevent
electromagnetic interference by maintaining a minimum distance
between portable and RF communications equipment (transmitters) and
HemoCue systems as recommended below, according to the maximum
output power of the communications equipment.
Rated
Separation distance according to frequency of
maximum
transmitter (m)
output power
150 kHz
80 MHz to
of transmitter
to 80 MHz
800 MHz
(W)
d=1.2√P
d=1.2√P
0.01
0.12
0.12
0.1
0.38
0.38
1
1.2
1.2
10
3.8
3.8
100
12
12
For transmitters rated at maximum output power not listed above, the
recommended separation distances (d) in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter,
where (P) is the maximum output power rating of the transmitter in
watts (W) according to transmitter manufacturer.
NOTE 1
®
Hb 301 Analyzer
NOTE 2
800 MHz to
2.5 GHz
d=2.3√P
0.23
0.73
2.3
7.3
23
39
At 80 MHz and 800 MHz, the separation distance for the
higher frequencyrange applies.
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.