Handpiece Trouble Shooting
Low Speed Handpiece
PROBLÈME
Cause
SOLUTION
Handpiece won't run
1. No or low air pressure.
2. Air line improperly sealed.
3. Inferior oil; too much or excess more commonly too little oil or water in the turbine.
4. Broken drive head.
Note new handpieces require a "Run in time" this can
be up to 20 hours of use
1. Check the air line to be sure the air is turned on and
the line is unobstructed.
2. Reconnect air line to handpiece making sure the
gasket is fully contracted.
3. See lubrication instructions on page 17.
4. Contact your local iM3 official Service Centre.
If the motor stops spinning place a fine screw driver
down the nose and rotate shaft.
Low Speed Handpiece - 4:1 Straight Nose Cone
PROBLÈME
Cause
SOLUTION
Cant fit R&R pro-phy angle
1. Plastic drive shaft broken off internally in Chuck housing.
2. Improper lubrication.
Only fit plastic iM3 R&R prophy angles Code L7465.
1. Contact your local official iM3 Service Centre.
2. See lubrication instruc-tions, on page 17.