7.
Accuracy Control
Remove the tube from the manometer and hold the gauge upright. Your
device is set correctly when the needle stops on the zero of the scale. If this is
not the case, the instrument must be recalibrated. Contact your medical
equipment reseller.
We recommend ta check every 2 years the accuracy of your device.
8.
Specifications
• Environmental conditions for which the minimum maximum tolerance of+ -
3mmHg is met:
- Measurement conditions 10 ° to 40 ° C, relative humidity of 85%, without
condensation
- Storage Conditions -20 ° to 70 ° C, relative humidity of 85%, without
condensation
• Display Type: Round scale
• Scale graduation: increment of 2 mmHg
• Display range 0 to 300 mmHg
• Measuring range 0 to 280 mmHg
• Mobility of the needle: not reset
• Pressure generation: bulb
• Pressure reduction: air-release valve.