6.2 Zero Calibration
6.3 Slope Calibration
7. Accessories
O
2
Changing it may only take place in case of repair, if the probe fails to respond or the head
is broken. (See chapter 6)
This head does not contain a lining or an electrolyte (optical functioning and not chemical).
8. After-Sale Services
The warranty is of 2 years.
For all settings, contact our Technical Support on 0 825 563 563.
The material must be returned to our workshops and for all repairs or spare parts, kindly
contact:
ENGLISH
-
Insert the sensor into a channel of an interface
On the sensor, go to « Eau » mode (meaning "water" mode) by clicking on the
-
orange button
-
Immerse the probe into a zero solution (saturated solution of anhydrous sodium
sulfite) during 2 to 5 minutes
-
Tap and hold the purple button for 10 seconds
-
Wait 30 seconds
-
Click briefly
-
Remove the probe from the solution and rinse it
-
The zero adjustment is finished.
-
Insert the sensor into a channel of an interface.
-
Leave the probe in the open air
-
On the sensor, go to « Air » mode by clicking on the orange button
-
On the sensor, tap and hold the purple button
-
Wait 10 seconds
-
Click briefly on the sensor
-
The slope adjustment is finished.