Data logging
Oxygen optical sensor with probe
Ref : 482 082
3.
Your probe comes ready to start.
Note: this setting can also be done in the software.
4.
4.1 Using the Probe in Standalone Mode:
The probe takes measurements in air and water. To go from one mode to another:
In the water, in case of outliers, an air bubble may be present under the probe head. To
remove it, tap on the probe body.
4.2 Using the Probe with the Scientific Workshop Software:
ENGLISH
First Use
-
Remove the probe protective cap.
-
Insert the probe into a channel of an interface
-
Wait a few seconds for the measurement to stabilize
-
If the value that appears on the sensor screen is different than 20.9 %, then
manually set the slope, as follows:
-
Tap and hold the purple button (3 seconds)
-
Once 20.9 % appears, press the purple button briefly
-
The slope of your probe is set.
Your probe is ready to operate
Measure
-
In standalone mode, press on the orange button
-
In the software, select « Water » in the sensor setting
-
Run the Standard Scientific Workshop software
9