Data logging
Oxygen optical sensor with probe
Ref : 482 082
You can control the setting by clicking on the « Measure » tab to notice that the value
measured by the probe corresponds to the O
The slope calibration can also be made directly in standalone mode, thanks to the sensor
screen.
See the detailed procedure in chapter 6.3 below.
5. Maintenance - Storage
Maintenance: After each use, rinse the probe head with distilled water.
Storage: Put the protective cap back. Keep the prob dry.
6. Changing the Probe Head
The probe head is designed to function 2 years. Beyond this period, it is preferable to
replace it with a new one (ref 453153).
6.1 Replacement
-
-
Once this operation is finished, a full calibration (zero adjustment + slope) of the
probe is necessary:
ENGLISH
You may now use your probe.
4.2.2
Slope Calibration in Standalone Mode
Unscrew the existing probe head
Screw the new probe head
rate in the Air, being 20.9 %.