Guidance and manufacturer's declaration – electromagnetic immunity
The Via Nano 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the Via Nano 6
should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the A.C. mains voltage prior to application of the test level.
T
Manuel Via Nano 6.1
IEC 60601 test level
+8kV contact
+15kV air
+2kV for power supply lines
+1kV line(s) to line (2)
+2kV line(s) to earth
0 % UT; 0,5 cycle
at 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
Compliance level
+8kV contact
+15kV air
+2kV for power supply lines
+1kV line(s) to line (2)
0 % UT; 0,5 cycle
at 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
30 A/m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the Via Nano 6 requires
continued operation during
power mains interruptions, it is
recommended that the Via
Nano 6 be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.