Display resolution
Automatic shutdown
Memory
Operational conditions
Battery
Weight & Dimension
SYMBOLS
Symbol
Description
Type BF applied part.
Information about a manufacturer
Please read the instructions carefully.
Waste electrical materials should be sent to a dedicated collection point for recycling.
SN
Serial number
LOT
Batch number
IMPORTANT
Inaccurate reading or thermometer damage may occur if the thermometer is not correctly used.
IP22
2 Protected against solid foreign objects of 12,5 mm Ø and greater;
2 If keep the thermometer in 15 degree angle, it still can prevent the water drop.
EMC INFORMATION
Guidance and manufacturer 's declaration-electromagnetic emissions
The infrared thermometer is intended for using in the electromagnetic environment specified below. The customer or
the user
of the infrared thermometer should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2 N/A
Voltage fluctuations
/flicker emissions IEC 61000-3-3
Guidance and manufacturer 's declaration-electromagnetic immunity
The infrared thermometer is intended for use in the electromagnetic environment specified below. The customer or the
user of the infrared thermometer should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst IEC
61000-4-4
NOT-AD - 70 Novembre 2020 (Version 1.0)
0.1℃/ F
10s±1s
35 groups of measured temperature.
Temperature: 10℃-40℃ (50 F-104 F) /
Humidity: 15-95%RH, non-condensing Atmospheric pressure: 86-106 kPa
2*AAA, can be used for more than 3000 times
66g (without battery),163.3×39.2×38.9mm
Compliance
Electromagnetic environment-guidance
Group 1
The infrared thermometer uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to
cause and interference in nearby electronic equipment.
Class B
The infrared thermometer is suitable for use in all establishments
other than domestic and those directly connected to the public low-
voltage power supply
N/A
network that supplies buildings used for domestic purposes.
IEC60601 test
Compliance
level
level
±2, ±4, ±6kV
±2, ±4, ±6kV
for Contact dis-
for Contact dis-
charge
charge
±2,±4,±8kV,
±2,±4,±8kV,
±15kV
±15kV
air discharge
air discharge
±2 kV for a.c.
N/A
power lines
±1 kV for d.c.
power lines
Electromagnetic environment -guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%
Mains power quality should be that of a typical commer-
cial or hospital environment.
17