English
Troubleshooting
No output signal
Maximum output signal
High setpoint vs. output offset
Signal lower than expected
Flow is gradually decreasing
Oscillation
Small flow occurs when actuator is
supposed to be closed
Parameter not available
Setpoint is neglected
Settings menu not accessible
Before switching on power, check if all connections have been made according to the hook-up diagram. The hook-
!
up diagrams are available at the download section of our website: http://www.bronkhorst.com
LED indications can be very useful in case of problems with the instrument. See the "MANUAL INTERFACE: micro-
switch and LED's" section in the manual "Operation instructions digital instruments (document nr. 9.17.023)".
This manual is available at the download section of our website: http://www.bronkhorst.com
Contact Bronkhorst® local sales representative or send e-mail describing your problem to:
support@bronkhorst.com
Doc nr.: 9.17.049
No power supply
PCB failure
Actuator blocked
Filter or screen blocked
Sensor failure
PCB failure
Sensor failure
PCB failure
Actuator blocked
Stoppage
Pressure not correct
System stoppage
(Gas) Condensation
Actuator adjustment has
changed
Controller adjustment wrong
Input pressure wrong
Actuator damaged
Volume / tubing between
controller and pressure
controller too small/ short
Wrong controller settings
Actuator is leaking
Parameter is not supported by
the connected instrument.
Instrument has wrong controller
mode
Alarm- or counterlimit reached
Security mode 'blocked' or
'menu'
Check power supply
Check cable
Return to factory
Clean actuator (qualified personnel only)
Clean filter or screen
Return instrument to factory
Return instrument to factory
Return instrument to factory
Return instrument to factory
Clean actuator (qualified personnel only)
Remove cause
Check pressure
Remove cause
Decrease supply pressure or increase
temperature
Check actuator adjustment
Check controller setting
Adjust pressure
Check actuator
Improve situation
Check controller settings
Check actuator
Set controller mode to 'digital input'
Reset alarm
Change security mode
8/16