Memory / memory space
Application part of type BF
Warning, observed accompanying leaflet
C A R E & C L E A N I N G
Cleaning
Clean the water-tight sensor every time after use with a soft cotton
which has been moistened with disinfectant. Clean the thermometer
casing with a soft, dry cloth.
The thermometer should not be washed off in water. On
no account should abrasive cleaners be used or the
casing of the forehead thermometer be immersed in
water or other liquids.
Protection
Do not expose the forehead thermometer to extreme temperatures,
high air humidity or direct sunlight. Avoid strong knocks or vibrations
and do not drop the thermometer. Do not touch the tip of the sensor
with the fingers. Do not take the forehead thermometer apart.
Fault finding and elimination
1. Temperature reading is always too low:
• The sensor was not drawn correctly over the forehead.
- The tip must be placed properly on the temple. If the
sensor is not properly positioned, this can cause
temperature readings to be too low.
• The water-tight lens is dirty.
- Clean the lens properly with a soft cotton cloth, which
has been moistened with some disinfectant.
• The lens is damaged.
- Consult your dealer or our customer service department.
2. The measured temperature is outside the measuring range.
• The scanning sequences were not carried out on the temple/
forehead.
- This may happen when the scanning sequence is started
before the forehead thermometer is placed correctly on
37