Télécharger Imprimer la page

SD BioSensor STANDARD M10 Fiche De Référence Rapide page 5

Flu/rsv/sars-cov-2
Masquer les pouces Voir aussi pour STANDARD M10:

Publicité

RESULT INTERPRETATION
EN
Result
Flu A Positive
Flu B Positive
RSV Positive
SARS-CoV-2 Positive
Negative
Invalid
- Each target (Flu A, Flu B, RSV, SARS-CoV-2) may have a co-infection.
- If the IC is negative and the results for any of the targets are positive, the results for all targets are considered valid. A high copy number of target-specific gene can lead
to reduced or absent IC.
- If an invalid result is confirmed in one or more of the pathogen results, that tests will be invalidated. Please conduct a re-test.
Outcome
Result
(Home screen)
(Review screen)
Positive
Negative
Invalid
X
Error
4
Flu
Flu A
Flu B
+
-
-
+
-
-
-
-
-
-
-
-
-
-
-
-
Description
At least one pathogen/
target gene is positive.
No pathogen was detected.
!
IC signal does not have a
Ct within the valid range.
The test failed because either
an error occurred or the test was
canceled by the user.
RSV
IC
RSV
+ / -
-
+ / -
-
+
+
+ / -
+
-
+
-
+
-
+
-
-
-
Outcome
Result
(Summary screen)
(Summary screen)
V
IC Valid
!
IC Invalid
X
IC Error
SARS-CoV-2
IC
ORF1ab
N
+
-
-
+
-
-
-
-
+
+
-
+
-
+
+
+
+
+
-
-
-
-
-
Description
IC has a Ct within the valid range.
: The test was completed. Report positive/negative results of pathogens
according to the interpretation shown in the console.
IC does not have a Ct within the valid range.
: Pathogens detected as positive shall be reported, but all negative
results (tested but not detected) are invalid. Repeat the test.
The test failed because either an error occurred or the test was canceled
by the user. Repeat the test.
IC
+
+
+
+ / -
+ / -
+ / -
+
-

Publicité

loading

Ce manuel est également adapté pour:

M10-cvfr-01