5. MAINTENANCE
For maintenance, use only specified spare parts. The
manufacturer will not be held responsible for any accident
occuring following a repair done other than by its After
Sales Service or approved repairers.
5.1
Servicing
When the
symbol is displayed, replace all the
batteries of the CA 6460 or recharge the battery of the CA
6462. Check that no other terminals are connected before
opening the equipment.
5.1.1
Replacing the batteries or rechargeable cells
(C.A 6460 only )
Detach the four captive screws under the housing
Then remove the shell + front panel assembly of the yellow
housing
Then undo the two screws closing the battery compartment
cover
Take out the 8 cells and replace them
Note: It is possible to replace the batteries by rechargeable cells
(1.2 V - 2 Ah or above, NiCd or NiMH, of same size):
Remove the plug under the batteries
Set the switch to the position: NiCd / NiMH
Replace the plug.
Insert the 8 rechargeable cells
Then in both cases:
Replace the battery cover
Undo the two screws closing the battery compartment
cover.
Replace the shell + front panel assembly in the yellow
housing then tighten the four captive screws under the
housing
5.1.2
Recharging all replacing the battery (C.A. 6462)
Connect the battery charge connector to the mains
The CHARGE indicator light comes on steadily in red
When the battery is charged, the CHARGE indicator light
comes on steadily in green
The charge time is approximately six hours for charging to
80% of the battery capacity. It is possible to top up the
charge.