PW2C... Single point load cells
1
Mounting and Load application
The load cells are fixed at the mounting bores. For the recommended screws
and tightening torques refer to the table below:
Max. capacity
7.2...36 kg
72 kg
*) Recommended value for the stated property class. For screw dimensioning please refer to the appropriate
information given by the screw manufacturers.
Load must not be applied to the side where the cable connection is located,
as this would cause a force shunt.
2
Electrical connection
The following can be connected for measurement signal conditioning:
S carrier-frequency amplifier
S DC amplifier
designed for strain gage measurement systems.
A2431-1.1 en/de/fr
Thread
M6
M6
platform center
load application
spacing washer
Min. property class
8.8
10.9
7
*)
Tightening torque
6 N@m
10 N@m
connection cable
mounting bores
mounting
HBM

Publicité

loading

Ce manuel est également adapté pour:

Pw2cc3Pw2cc3mrPw2cc6Pw2cc6mr