Determination of free chlorine HR with powder packs
Select the method on the device.
5 mL
10 ml
3
5 ml
Fill 10 mm vial with 5 mL
sample.
Zero
Press the ZERO button.
10 ml
3
5 ml
Close vial(s).
EN   Method Reference Book   1.0
10 ml
3
5 ml
Close vial(s).
Remove vial from the
sample chamber.
Invert several times to mix
the contents (20 sec.).
Chlorine HR PP / M111
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
10 ml
3
5 ml
Add two Chlorine FREE-
DPD / F10 powder packs .
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
EN
13