KS4.3 T / 20
Method name
EN
Measuring range
Chemical Method
4
How to use
K
T
S4.3
0.1 - 4 mmol/l K
S4.3
Acid / Indicator
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type
Cuvette
MD 200, MD 600, MD 610,
ø 24 mm
MD 640, MultiDirect, PM 620,
PM 630
SpectroDirect, XD 7000,
ø 24 mm
XD 7500
Material
Required material (partly optional):
Reagents
Alka-M-Photometer
Alka-M-Photometer
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to K
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Language codes
Method number
λ
Measuring Range
610 nm
0.1 - 4 mmol/l K
615 nm
0.1 - 4 mmol/l K
Packaging Unit
Part Number
Tablet / 100
513210BT
Tablet / 250
513211BT
S4.3
ISO 639-1
NL Handboek van Methoden 01/20
EN Handbook of Methods 01/20
EN   Handbook of Methods    
Bar code for the
detection of the
methods
20
S:4.3
Display in the MD
100 / MD 110 /
MD 200
S4.3
S4.3
are iden-