14
РМ 2522
V. Installation
DIRECTIONS FOR USE
V-1.
EARTHING
The instrument must be connected to a wall socket with rim earthing via the 3-pole mains
cable supplied on delivery.
|
The housing of the instrument is then earthed via the above-mentioned cable.
If the instrument cannot be earthed via the mains cable, a separate earthing cable should be
connected to the earthing socket (13) at the front of the instrument (see Fig. 11, page 52).
Pay attention to the local safety directions.
V-2.
MAINS SUPPLY
The instrument is set to 220 V in the factory.
By connecting the transformer windings as shown in Fig. 9 , page 48, the instrument can be
used at the following voltages.
а.
220 V + 10%; -15%
fuse
100 mA slow blow
b.
ПОМ
+ 1096; -15%
fuse
200
mA slow blow
V-2.1.Battery supply
The optional accessory PM 9216 is recommended for battery supply, as it becomes an integral
part of the instrument.
V-3.
CONTROLS AND CONNECTORS
Front panel (Fig. 11, page 52)
POWER
Switches on the instrument
V===; Vou; mA==; | Switches on the required measuring function
тА;
КО; 20 MQ
HOLD
O2 220;
Display hold; Range selector of all functions
200; 2000
except 20 МО range
Combined Hi-input of voltage and resistance
measurements
Hi-input terminal of current measurements
Lo-input terminal
Earthing terminal
Zero adjust