3.
SPECIFICATIONS
Measuring ranges
Output sensitivity
Accuracy (45-65Hz)
Minimum load
Frequency range (-2dB) 20Hz to 20kHz (ACP 1000)
Phase (45-65Hz)
Position sensitivity
External field
Noise (ACP 1000)
Noise (ACP 1005)
Temperature coeff.
Power supply
(ACP 1005 only)
Battery life
External Power
Low battery
Overload
Enclosure
Material
Degree of Protection
Dimensions
Output (ACP 1005)
Output (ACP 1000)
_______________________________________________________________
5
10A / 100A / 1000A AC
100mV / 10mV / 1mV per A (AC coupled)
± 1% of reading ± 0.1A 10/100A ranges
± 1% of reading ± 1A
100kΩ for specified accuracy
20Hz to 100kHz (ACP 1005)
< ± 1°
± 2.5 % of reading
± 0.25% of range with cable >200mm (8")
from the probe
0.08A (10A/100A/1000A range)
0.03A / 0.075A / 0.075A (10A/100A/1000A range)
± 0.15% of reading/°C
Two AA MN1500 LR6 alkaline batteries
or an external power supply (+3V/100mA)
1000 hours typical ACP 1000
150 hours typical ACP 1005
Via 2.1mm (0.08") mini power connector
Indicated by an orange LED
Indicated by a red LED
ARNITE, UL94 V0
IP40 (not certified by UL)
110 (l) x 65 (w) x 23 (d) mm
0.5m coax cable terminated with a safety BNC or
0.5m coax cable terminated with a safety 4mm plugs
1000A range