Product HANDBOOK
Special Instructions for medical applications (IEC/EN/UL/CSA 60601-1)
These products are designed for continuous operation within an overall enclosure, and must be mounted
such that access to the mains terminals is restricted. See clause 16, IEC60601-1, EN60601-1 and UL/CSA
60601-1.
Input Connection - KMx15 –xx
Pin
Single O/P
1
No pin
2
AC (L)
3
AC (N)
4
-DC out
5
No pin
6
No pin
7
+DC out
8
No pin
9
No pin
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with
oxygen, or with nitrous oxide.
These products are classed as ordinary equipment, type B and have a reinforced barrier between input and
output to IEC60601-1, EN60601-1 and UL/CSA 60601-1. They are NOT protected against the ingress of
water.
Reference should be made to local regulations concerning the disposal of these products at the end of their
useful life.
Environmental Specifications:
Description
Use
Temperature
Humidity
Altitude
Pressure
Orientation
Shock
Vibration
Material Group
Pollution Degree
Overvoltage Category
Class
KM15 Series
Dual O/P
Triple O/P
No Pin
No Pin
AC (L)
AC (L)
AC (N)
AC (N)
-DC out
GND for +5V
GND
+5V
No pin
-DC out
+DC out
No pin
No pin
GND
No pin
+DC out
Operation
Indoor
-25° C - +70° C (derating at 2.5%/° C above 50° C)
20 - 95% RH, non-condensing
-200m - 3000m
70kPa - 106kPa
2
23.52m/s
(10-55Hz): constant sweep 1min X, Y, Z for 1 hour
2
<196.1m/s
(20G)
IIIb
2
II
II
Storage
-
-40° C - +100°C
20 - 95% RH, non-
condensing
-200m - 5000m
54kPa - 106kPa
ALL