16
Error Messages
Error Message
Hi /Lo
Er2
Er3
Er 5 to 9
Error
The temperature being
measured is outside of the
settings of HAL and LAL.
The temperature being
measured is outside of the
measurement range of the
meter.
The meter is exposed to
rapid changes in the ambi-
ent temperature.
T
< 0°C (32°F) or
ambient
T
> +50°C (122°F)
ambient
Measure
The meter should be allowed plenty of time
(minimum 30 minutes) to stabilize to the
working/room temperature.
Reset: Turn the meter off, remove the bat-
tery and wait for 1 minute, reinsert the bat-
tery and turn on.
Send the device in for repair if the error
message still appears after the reset.
Error Messages
45
EN