Optima Turn / English
(5). Modes
A. Alternate Mode
The air cells inside the mattress are deflated and inflated periodically to avoid prolonged
pressure in certain areas of patient.
B. Continuous Low Pressure
Function LED indicator is steady-on in constantly non-alternating mode. The system will
redistribute body mass over a greater surface area at a constant low pressure.
C. Seat Inflation
Press the
supports to the patient during upright position without bottoming out.
(6). Turning Mode
Three turning modes including
TURN can be selected. Turning angle is from patients' direction and hold at that position for dwell time.
Top air cells are non-alternating during the turning process, and will change to alternating mode when it
reaches the position of the desired angle and when it comes back to supine position as well.
NOTE- When dwell time of 10 min is selected, even reaches the position of the
desired angle it doesn't change to alternating mode.
(7). Turning Angle
45.7cm)
Light-turn,
⚫ For One-Time Turning which indicated with LED flashing, press and hold
seconds. When turning is completed, it will go back to supine position with alternating mode.
(8). Pressure Adjust Levels
⚫ Press the
⚫ The mattress pressure can be manually adjusted for patient comfort using the key
firmer setting and decrease
NOTE- When the power is started (i.e., the surface is inflated), it will automatically go
to "Max Firm" mode to hasten inflation. Once the system is ready, the system will
enter CLP mode for 10 minutes. If there are no changes of modes, system will switch
to Alternate Mode.
NOTE- You can also check if the pressure is suitable for the patient by sliding one
hand beneath the air cells at the level of the patient's buttocks. Always leave at least 1"
inch space between patient and the static cell to prevent bottoming out.
key to enter the desired mode in alternate or CLP mode for additional
LEFT TURN,
(Up to 30 degree turning under Full-turn Mode, subject 75Kg with shoulder width
Half-turn,
/
to get the correct setting of the pressure level based on patient weight.
for a softer setting.
RIGHT TUTN and
Full-turn
LEFT-SUPINE-RIGHT
key for 3
for a