4. USB Type-B: Connect the camera to a desktop or laptop for video transmission when using
third-party video conferencing software such as Skype or Teams (Note).
5. HDMI: Connect the camera to a TV or a monitor to display video output. The camera and the
connected TV or monitor must have grounding design. (Note).
6. Power: The camera and the connected TV or monitor must have grounding design. Use the
supplied power adapter and power cord to connect the camera to a power outlet and make sure
the power cord of the TV or monitor supports the grounding plug.
Note:
Accessory equipment connected to the analog and digital interfaces must be in compliance with the
respective nationally harmonized IEC standards (i.e. IEC 60950 for data processing equipment, IEC
60065 for video equipment, IEC 61010-1 for laboratory equipment, and IEC 60601-1 for medical
equipment.) Furthermore all configurations shall comply with the requirements of the system in
standard IEC 60601-1. Everybody who connects additional equipment to the signal input part or signal
output part configures a medical system, and is therefore, responsible that the system complies with
the requirements of the system in standard IEC 60601-1. The unit is for exclusive interconnection with
IEC 60601-1 certified equipment in the patient environment and IEC 60XXX certified equipment
outside of the patient environment. If in doubt, consult the technical services department or your local
representative.