5 Replace the white calibration tile holder.
6 Reconnect the cables.
Clean the internal feeding mechanism
Clean the internal feeding mechanism as needed.
Do not clean any parts of the instrument other than the ones that are specified.
Do not use compressed air to blow dust out.
Do not get dust on the white calibration tile.
1 Turn off the ES-6000.
2 Disconnect the cables from the ES-6000.
3 Insert a flat screwdriver in the slot at the center of the cover and turn it to open the cover.
4 Move the measuring head carriage all the way to the left by pushing the plastic housing of the motor. Touch only
the motor.
5 As you manually turn the roller wheel, use water or alcohol and a lint-free wipe to clean the bottom media guide
assembly and the rubber rollers.
6 Close the cover and reconnect the cables.
Prepare for transport
If you transport the ES-6000 over a long distance, prepare the instrument first.
1 Turn off the ES-6000.
2 Disconnect the cables from the ES-6000.
3 Set the transport guard by moving the switch on the bottom of the instrument to the locked position.
4 Insert a flat screwdriver in the slot at the center of the cover and turn it to open the cover.
5 Move the measuring head carriage all the way to the right by pushing the plastic housing of the motor, until the
lock engages. Touch only the motor.
6 Close the cover.
Safety
Follow safety guidelines to protect the ES-6000.
The ES-6000 is classified as a class 1 LED product in accordance with safety standard EN/IEC60825-1. To avoid any
risk of contact with the visible and invisible light from the LED, do not operate the instrument with the cover open
or the white calibration tile open.