All manuals and user guides at all-guides.com
Accuracy
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 80% relative humidity.
AC Current
Range
Accuracy
199.9 µA
± (1.5% + 0.4 µA)
± (1.5% + 0.004 mA)
1.999 mA
± (1.5% + 0.04 mA)
19.99 mA
± (1.5% + 0.4 mA)
199.9 mA
± (2.5% + .04 A)
10.00 A*
* Duty Cycle: 60 seconds on, 5 minutes off
DC Current
Range
Accuracy
199.9 µA
± (1.2% + 0.1 µA)
± (1.2% + 0.001 mA)
1.999 mA
± (1.2% + 0.01 mA)
19.99 mA
± (1.2% + 0.1 mA)
199.9 mA
± (2.0% + 0.03 A)
10.00 A*
* Duty Cycle: 60 seconds on, 5 minutes off
AC Voltage
Range
Accuracy
± (1.2% + 0.4 mV)
199.9 mV
± (1.2% + 0.004 V)
1.999 V
± (1.2% + 0.04 V)
19.99 V
± (1.2% + 0.4 V)
199.9 V
± (1.5% + 4 V)
600 V
DC Voltage
Range
Accuracy
± (0.5% + 0.1 mV)
199.9 mV
± (0.8% + 0.001 V)
1.999 V
± (0.8% + 0.01 V)
19.99 V
± (0.8% + 0.1 V)
199.9 V
± (0.8% + 1 V)
600 V
Frequency Range
Burden Voltage
< 0.7 V
40 Hz to 500 Hz
< 0.3 V
Burden Voltage
< 0.7 V
< 0.3 V
Frequency Range
Input Impedance
40 Hz to 500 Hz
10 MΩ
Input Impedance
10 MΩ
93-606
13