OPERATING PRECAUTIONS
■ When using a D clamp first connect it to the measuring instrument before clamping
around the conductor.
■ Do not use the clamp when the voltage, with respect to earth, of uninsulated busbars or
conductors exceeds 600V rms sinusoidal or 848V crest.
■ Never leave the clamp around a conductor unless it is connected to a measurement
instrument.
■ Remove the clamp from the conductor before changing range.
■ Maintain the cable or busbar central in the clamp and hold it perpendicular to the
conductor.
■ Whenever possible avoid the proximity of other conductors which might produce fields
of interference.
■ Keep the jaw faces clean. If necessary clean them with a lightly oiled soft rag, to avoid
oxydisation.
■ Avoid leaving the clamp in very damp places or exposed to running water.
Clamp D30N LA 3000/1.....................................................................1200.49A
Clamp D30CN LCA 3000/1..............................................................1200.64A
Clamp D31N LA 500-1000-1500/1..................................................1200.50A
Clamp D32N LA 1000-2000-3000/1...............................................1200.51A
Clamp D33N A 3000/5........................................................................1200.52A
Clamp D34N A 500-1000-2000-3000/5.........................................1200.53A
Clamp D35N A 1000-2000-3000/5..................................................1200.54A
Clamp D36N LA 3000/3.....................................................................1200.55A
ACCESSORIES
For all models
- Lead n° 153A/153B (pair of 1metre leads).......................................1007.41
For model
D36- Lead n°163 (1,5m for connection to NUWATT and NUPHI)...............1007.81