351 / 352 SPECIFICATIONS
TEMPERATURE RANGES:
35100-K / 35200-K: -100° to 999°F (-73° to 537°C)
35100-J / 35200-J: -100° to 1382°F (-73° to 750°C)
35100-T / 35200-T: -100° to 752°F (-73° to 400°C)
35132 / 35135 / 35140: -100° to 500°F (-73° to 260°C)
35232 / 35235 / 35240: -100° to 500°F (-73° to 260°C)
ACCURACY:
35100-J,K,T / 35200-J,K,T:
With factory calibration, ±0.5°F (±0.3°C) over entire
measurement range at ambient temperatures between
68° to 86°F (20° to 30°C)
35132 / 35140 / 35232 / 35240:
With factory calibration, ±0.9°F (±0.5°C) Total System
Accuracy over entire measurement range at ambient
temperatures between 68° to 86°F (20° to 30°C)
35135 / 35235:
Instrument Only: With factory calibration, ±0.5°F
(±0.3°C) over entire measurement range at ambient
temperatures between 68° to 86°F (20° to 30°C).
Total System Accuracy: Surface probe temperature
RESOLUTION:
35100-J,K,T / 35132 / 35135 / 35140:
Tenth degree (0.1)
35200-J,K,T / 35232 / 35235 / 35240:
Tenth degree / whole degree (0.1/1), user selectable
BATTERY LIFE:
1800 Hours (2) AAA Alkaline batteries
AMBIENT TEMPERATURE LIMITS:
32° to 122°F (0° to 50°C)
TRACEABILITY:
We certify that this instrument has been calibrated to
temperature and/or electrical standards traceable
to the U.S. Department of Commerce National Institute
of Standards and Technology.
CALIBRATION:
Under normal operating conditions, the unit should not
require recalibration.
errors for draft-free, clean
oiled surfaces with 2.2 lbs
(1 Kg) of pressure are
typically within +2.7°F and
-6.3°F (+1.5°C and -3.5°C)
without thermostat recycling.