Heating Drying Oven
Oven 50 litres
Ref :
535 017
ENGLISH
Procedure of « temperature setting » and « time setting » :
Power on
PV window : actual temperature
SV window : last setting temperature
Press « SET » key once to enter
« temperature setting » status
PV window : actual temperature
SV window : last setting temperature flashing
Use « shift key », « up key » and « down key » to set
required setting temperature value
Remark : Except the setting of temperature, all the other data of setting have
been locked. If you wish to revise the new data. First cancel the command for
locking out modification of the default parameters.
6
Operation
1. Put the sample into the chamber. Close the door and turn the 'Exhaust flap'
to 'Z' position.
2. Turn the power switch to « 1 » .The indicator of the power supplier will turn
on and the temperature will be digitally displayed on the controller.
3. Temperature setting
After having switched on the unit, if the set temperature is the same as the
needed working temperature, then it is not necessary to set it again.
Otherwise, it is necessary to reset the working temperature. First, press the
function key 'SET' on the control panel to enter the temperature setting
program, till the setting temperature display SV is flashing. Then press the
'SHIFTING KEY' and the 'INCREASING KEY' or the 'DECREASING KEY' to
set the needed working temperature. Finally, press again the' SET' key for 5
seconds to confirm the setting parameters.
For example, the needed working temperature is 150.0°C, and the former
setting temperature is only 86.5°C. First, press the function key 'SET' on the
control panel to enter the temperature setting program, till the setting
9
Press « SET » key for 5 seconds
Press « SET » again to confirm
setting values and to exit setting
procedure
Control Parameters Setting
AL1 : over-temperature
alarm setting value
AL2 :
SC : modification of
temperature display
ATU : self-adjust function
P
I
D
T
LCK
See remark