Télécharger Imprimer la page

IFM Electronic efector500 PF20 Notice Utilisateurs page 31

Capteur de pression électronique

Publicité

Les langues disponibles

Les langues disponibles

Voltage output 0 ... 10 V
Factory preset
U [V]
10
-1
0
MEW = final value of the measuring range
The output signal is between 0 and 10 V in the set measuring range.
It is also indicated:
System pressure above the measuring range: output signal > 10 V.
Current output 4 ... 20 mA
Factory preset
I [mA]
20
4
-1
0 (ASP)
MEW = final value of the measuring range
The output signal is between 4 and 20 mA in the set measuring range.
It is also indicated:
• System pressure above the measuring range: output signal > 20 mA.
• System pressure below the measuring range: output signal drops to max.
3.2 mA (depending on the scaling).
32
P
MEW
P
MEW (AEP)
Measuring range scaled
U [V]
10
-1
0
ASP
Measuring range scaled
I [mA]
20
4
-1
0
ASP
P
AEP
MEW
P
AEP
MEW

Publicité

loading