Parker 20H Guide D'utilisation page 27

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 87
No.
C o n d u c t
14
A C T
1 0 0 P S I
P r e s s u r e
15
A C T
1 0 0 P S I
W a t e r
P u m p
16
A C T
1 0 0 P S I
C a l
i b r a t i o n
17
A C T
1 0 0 P S I
W a t c h d o
18
A C T
1 0 0 P S I
F L O A T
T R A N S
19
A C T
1
0
0
P S I
F L O A T
C H A M
20
A C T
1 0 0 P S I
S e r v i c e
21
A C T
1 0 0 P S I
O
V e n t
B l o c k e d
2
22
A C T
1 0 0 P S I
R e m o t e
23
A C T
1 0 0 P S I
L o
c
k
24
A C T
1 0 0 P S I
Error Message
T r a n s
There is a fault with the conductivity sensor.
T r a n s
There is a fault with the pressure transducer
or wiring.
There is a fault with the water pump.
The calibration file is out of date.
g
Internal software time out.
There is a fault with the float chamber water
level transducer.
B
E R
There is a fault in the float chamber; the
water level is too high.
R e q
d
A service is required.
The oxygen vent is blocked.
S t o p
The terminals on JP3 of the options board
are short circuit.
Pressure error, O
consecutive errors of the same type have
occurred.
H2 Production
(System locked)
vent blocked or three
2
(System locked)
Action
No
Contact Parker Hannifin for advice.
Yes
Perform the required service.
Remove the blockage and perform a
hard reset.
No
If the problem persists contact Parker
Hannifin for advice.
Press
when required to reset
No
the generator.
No
Resolve the initial fault and reset the
generator.
23