For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
The PO3M is intended for use in the electromagnetic environment specified below. The
customer or the user of the PO3M should assure that it is used in such an environment.
IMMUNITY
test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC/EN 61000-4-8
NOTE: UT is the A.C. mains voltage prior to application of the test level.
All manuals and user guides at all-guides.com
Compliance
IEC 60601test
level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment
- guidance
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the
relative humidity should be at
least 30 %.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.