OPERATING INSTRUCTION
AC Current Measurement
1)
Make sure the "Data Hold" switch is not pressed.
2)
Set the Function / Range Switch to the required ACA range.
3)
Press the trigger to open the transformer jaws and clamp
one conductor only. It is impossible to make measurements
when two or three conductors are clamped at the same time.
4)
The value displayed on the LCD is the AC current flowing
through the conductor.
Insulation Resistance Test
1)
Set the Function / Range Switch to the insulation tester
2000MΩ range. In this condition, it is normal that the reading
is unstable.
2)
Insert the three banana plugs V.Ω, COM, EXT from the
insulation tester unit to the corresponding V.Ω, COM, EXT
input jacks on the clamp meter.
3)
Set the Function / Range switch of the insulation tester unit
to the 2000MΩ position.
4)
Connect the test leads from the insulation tester unit to the
appliance to be tested (the LED on the insulation tester unit
will light).
5)
Set the insulation tester Power switch to the "ON" position.
6)
Push the "PUSH-500V" button, the red LED "500V" will light.
The reading on the LCD of the clamp meter is the insulation
resistance value; if the reading is below 19MΩ, set the rotary
switch to range 200MΩ to increase the accuracy of the
measurement.
7)
If the insulation tester unit is not used, the power switch must
set to OFF position. And the test leads must be removed
from the input jacks; this can extend the battery life and
prevent electrical shock hazard.