Instruction Manual
The Clean and Drybox cleans the hearing aid by
means of UVC and dries by means of heating. Hereby
the moisture is removed by the heating cycle. The
UVC light removes dirt and the possible presence of
bacteria from your hearing aids.
This allows the hearing aid to perform better,
so that it creates a better hearing effect and can help
to prevent skin irritations, itching, rashes or ear
infections.
WARNING:
THE CLEAN & DRY BOX 5.0 CONTAINS HIGH
PRESSURE. IT SHOULD ONLY BE OPENED
OR REPAIRED BY PROFESSIONALS.
SUITABLE FOR RECHARGEABLE HEARING
EQUIPMENT
IMMEDIATELY REPORT ANY SERIOUS
INCIDENTS TO THE MANUFACTURER.
Instructions for use:
1. Please insert one end of the supplied MICRO
USB cable into the MICRO USB port of the
Clean & Dry Box 5.0.
2. Insert the other end of the cable into the USB
port of the transformer. After connecting it to
the power grid, five LEDs on the Clean & Dry
Box 5.0 will light up from left to right. This will
be repeated three times, at which point the
Dry Box is now ready for use. The standby
function is now enabled.
3. Clean the hearing aid according to the
manufacturer's cleaning instructions. After
wiping it with a dry towel, place the hearing
aid inside the Clean & Dry Box 5.0, distributed
across the right and left side of the tray and
close the lid. For a non-rechargeable hearing
aid, remove the batteries from the device.
4. Press the start button once. The five LEDs will
light up (four white and one blue) and the
Clean & Dry Box 5.0 will begin its 5-hour cycle.
5. When all of the LEDs are turned on, the warm
air flow starts to work. An airflow of 30
degrees Celsius flows along the sides of the
box to warm up the hearing aid during this
period. Simultaneously, the UV-C light
(wavelength 254 nm) is turned on, which
cleans the hearing aid. After approximately
eight minutes, the cleaning process is
complete and the UV-C light turns off,
however the flow of hot air continues.
The LEDs indicate the remaining heating
period:
First 75 minutes
Second 75 minutes
Third 75 minutes
Fourth 75 minutes
Finally, LED 5 will turn green and start to blink.
The entire 5-hour cycle is now complete. The
flow of hot air will be switched off. If the
device is accidentally opened while the UV-C
light is still on, the computer will turn off the
UV-C light safely and end the calculated UV-C
time. After completion of the heating period,
LED 1 blinks and
then turns off
LED 2 blinks and
then turns off
LED 3 blinks and
then turns off
LED 4 blinks and
then turns off
the Clean & Dry Box 5.0 will turn off
automatically and the green LED will continue
to blink for 3 hours. If you have to interrupt
the dehumidification process or the UV-C
process, press and hold the start button for at
least 3 seconds or disconnect the power
source.
6. Let your hearing aid cool down for at least 5
minutes, then the device is ready for use
again.
7. When using a different transformer, the
delivered Voltage should be between 4.75V
and 5.25V, and the current should be
approximately 500mA, otherwise the Clean &
Dry Box 5.0 may break down or it may not
work properly.
8. The Clean & Dry Box 5.0 automatically
recognizes the USB2.0 computer current
(500mA); you only need to connect it to the
USB port. If the USB power source only
supplies 500mA, the Clean & Dry Box 5.0 will
be set to 500mA mode. Because the heating
power is small, the temperature might not
reach 30 degrees Celsius. If the USB plug is not
inserted correctly or the effectively supplied
current is below 500mA, the Clean & Dry Box
5.0 will stop working.
When you want to leave the 500mA mode,
please disconnect the transformer for at least
10 seconds. Reconnect the device with a
transformer that supplies a current of more
than 1A, and a Voltage of 5V. The Clean & Dry
Box 5.0 will restart in 1000mA mode.
9. When the device is not in use for a longer
period of time, or when leaving the house,
please disconnect the device from the power
source.
Care of the UVC Clean & Drybox 5.0:
When cleaning your UVC Clean & Dry Box 5.0,
first disconnect it from the power supply. We
recommend using moist wipes for cleaning the
drying station. Please avoid direct contact with
water. Only the inner element can be removed
and may be rinsed under running water (see
illustration on page 4).
Specifications:
Model TC-CDB-19
Input 4.75-5.25V
UVC light 1.5W +-10%
Heating 3.5W +-10%
Heating time 5 hours
UVC time 8 minutes
Adaptor input 100 V- 240 V;50 HZ /
Output 5 V d.c.; 1000 mA
Working conditions Indoor use
Operation altitude
range
Operating
temperature
Supply voltage
fluctuations
Pollution degree Class II
60 HZ; 0.2 A
0 m up to 2000 m
5 degree Celsius to
40 degree Celsius
Not more than 10 %