Temperature Measurements; Mv Measurements - Sartorius PT-15 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Temperature Measurements

The meter can be used to measure
temperature independently with the
ATC/Temp sensor, without using
the pH sensor.
1.
Place the ATC/Temp sensor in the
media to be measured. The measured
temperature is displayed.
2.
If "----" is displayed, the temperature
is out of the temperature range –10 to
120°C.
Once the temperature is within range,
then "----" will disappear and the
correct temperature reading will be
displayed.
12

mV Measurements

1.
Measurement in the mV mode:
1.1 Connect an ORP sensor to the BNC
connector of instrument.
1.2 Press the [MODE] key until the LCD
displays the mV icon.
1.3 Rinse the sensor in distilled water.
Then rinse with a little of the sample
(if enough is available) and immerse in
the sample.
1.4 Allow sufficient time for the display
to stabilize. The instrument will
display the mV value of the sample.
If "----" is displayed, the mV value
measured is out of the "–1999/
+1999 mV" range. The instrument will
display the correct value once the
input mV is brought within range.