MAINTENANCE

The leak detector requires maintenance when used regularly:

Replace Filter:

Unscrew the sensor tip, as shown below. Replace the filter whenever it
becomes visibly dirty or every 3 to 4 months, depending on use.
Cleaning:
Clean the screen and the housing surface with a cloth and cleaning agent.
Do not use corrosive cleaning agents or solvents to clean the unit. Weak
household cleaners and soap suds may be used.
8.0

TROUBLESHOOT

Cross Sensitivity to Automotive Chemicals
Some automotive solvents and chemicals have similar hydrocarbon
properties as R134a and may elicit a positive response. Before leak
checking, remove any chemicals in the list below that elicit a positive
response.
Chemical Name/Brand
Rain-X Windshield Wash Fluid
Ford Sport Remover (Wet)
Ford Rust Inhibitor
Ford Gasket Adhesive (Wet)
Loctite Natural Blue degreaser (diluted)
Ford Brake Parts Cleaner
Ford Silicone Rubber (uncured)
11
Response
Yes
Yes
Yes
Yes
Yes
Yes
Yes