Télécharger Imprimer la page

Medline MDS3003 Guide D'utilisation page 10

Publicité

Les langues disponibles

Les langues disponibles

10. Information regarding potential electromagnetic or other interference between the
blood pressure monitor and other devices together with advice regarding avoidance
of such interference please see part ELECTROMAGNETIC COMPATIBILITY
INFORMATION.
11. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the
procedure of blood pressure measurement, a signal of
Under this condition, the Electronic Sphygmomanometer can keep function,
but the results may not be accurate; it's suggested that you consult with your
physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s, and the number of such pulse
takes more than 53 percent of the total number of pulse.
12. Please do not use the cuff other than supplied by the manufacturer; otherwise it
may bring biocompatible hazard and might result in a measurement error.
13.
The monitor might not meet its performance specifications or cause safety
hazard if stored or used outside the specified temperature and humidity ranges in
specifications.
-10-
 
will be displayed.

Publicité

loading

Ce manuel est également adapté pour:

Mds3003p