eng3
16 Compatibilité électromagnétique
Émissions électromagnétiques
The NanoVi is intended for use in the electromagnetic environment specified below. The customer or the user of the NanoVi should assure that it is used in
such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Immunité électromagnétique
The NanoVi is intended for use in the electromagnetic environment specified below. The customer or the user of the NanoVi should assure that it is used in
such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
lEC 60601 test
level
± 8 kV contact
± 8 kV contact
± 2 kV air
± 2 kV air
± 4 kV air
± 4 kV air
± 8 kV air
± 8 kV air
± 15 kV air
± 15 kV air
± 2 kV for power supply
± 2 kV for power supply lines
lines
Repetition frequency 100 kHz
± 1 kV for SIP/SOP
Repetition frequency
100 kHz
± 0,5, 1,0 kV line(s) to
± 0,5, 1,0 kV line(s) to
line(s)
line(s)
± 0,5, 1,0, 2 kV line(s) to
± 0,5, 1,0, 2 kV line(s) to
earth
earth
Electromagnetic environment – guidance
The NanoVi uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
The NanoVi is suitable for use in all establishments, including domestic and medical
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance level
Electromagnetic environment - guidance
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
25