Dilution
If the peroxyacetic acid concentration being measured is
above 2000 mg/L, perform a dilution.
1
Fill the dilution tube to the line labelled with the
dilution factor.
2
Fill the tube with deionised water to the line
marked "deionised water"
3
Cap the tube and mix until combined.
4
Fill the front part of the PAASense sample vessel to
the brim with diluted solution. Proceed with testing
as normal.
Discard the remainder from the dilution tube, and
rinse with deionised water.
5
Multiply the value displayed on screen by 2.
Record the product as the test result.
For example:
Displayed on the screen: 1750 mg/L
Dilution factor: 2
Test result: 3500 mg/L