6. Operation

6.1 Insertion of the capsule

Open and fold back the cover of the CapMix device. Press
down on the release button to release the capsule holder.
Keep the release button depressed and push to the left to
open the capsule holder.
Place the activated capsule between the two cups of
the holding mechanism and release the release button.
The device clamps the capsule automatically. Insert
capsules with tips such that the tip points towards the left
side.
Do not operate the device without a capsule inserted!

6.2 Mixing

Close the protective cover and press the "Start/Stop"-button.
This starts-up the mixing process.The countdown of the
mixing time is shown on the time display.
After the mixing time has elapsed, the latest time setting
entered is re-displayed so that there is no need to re-enter
any value if the same time is to be used.
The selected program can be interrupted at any time by
pressing the "Start/Stop"-button, upon which the program
returns to its original setting.
CAUTION: For safety reasons, the unit cannot be started
up unless the cover is closed. Attempting to operate the
unit with the cover being open causes the time display to
flash and an error message to be issued. Opening of the
protective cover during mixing is associated with a high
risk of injury. Hence, the mixing process is aborted.This
may render the capsule useless! Therefore, never open
the cover before the mixing process is completed and the
capsule holder has come to a complete stop!

6.3 Removal of the capsule

Open the lid of the CapMix unit.
Open the capsule holder by pressing the release button
down and towards the left, and remove the capsule.

7. Notes

Capsule materials of third-party manufacturers (including
amalgam capsules) may be mixed with the CapMix device
provided they are suitable for the device in terms of weight
and dimensions (please refer to the section, Technical Data).
The optimal mixing time for mixing with the CapMix device
should be inquired from the capsule manufacturer.
Pay attention to capsule leak tightness when mixing.
Capsules that are not leak tight contaminate the device
with mercury.

8. Operating Errors

Error
Possible
cause
Time display is off,
Power failure
even though the
unit is switched on
Motor fails to run
Cover of the
when the unit is
unit is open
switched on
Motor is
overheated
NOTE: The CapMix device should be repaired by 3M or an
3M-authorized repair facility only.

9. Maintenance and Care

The device requires no maintenance.

9.1 Fuse Replacement

Shut the unit off and disconnect the power cable. The fuses
are located on the back of the unit next to the power switch.
With a screw driver press down on the small bracket of the
fuse unit, squeezing the unit out a small distance.
Fuse unit
Remove the fuse unit from the device.
Remedy
Check both the
power cable and
main voltage
Close protective
cover
Allow the unit
to cool
3