Calibration Mode
GB
Store
Date
Cal
4
Time
Cal
Note:
user calibration (Display in calibration mode)
cAL
CAL
factory calibration (Display in calibration mode)
After confirming the selection with the [MODE] key the instrument
CAL
will show CAL/"Method".
METHOD
Scroll through methods using the [MODE] key.
Fill a clean vial with the standard up to the 10 ml mark, screw the
cap on and place the vial in the sample chamber making sure that
the
Zero
Press the [ZERO/TEST] key.
Test
METHOD
The method symbol flashes for approx. 8 seconds.
0.0.0
The display shows the following in alternating mode:
CAL
Perform calibration with a standard of known concentration (see
"Operation").
Zero
Press the [ZERO/TEST] key.
Test
METHOD
The method symbol flashes for approx. 3 seconds.
RESULT
The result is shown in the display, alternating with CAL.
CAL
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Mode
Pressing the [ZERO/TEST] key once decreases the displayed value by
Zero
Test
1 digit.
CAL
Press the corresponding key until the reading equals the value of the
calibration standard.
RESULT + x
By pressing the [ON/OFF] key, the new correction factor is calculated
On
and stored in the user calibration software.
Off
:
:
Confirmation of calibration (3 seconds).
30
All manuals and user guides at all-guides.com
User calibration
4
marks are aligned.
MD200_1 09/2011