Per Iec1010 Overvoltage Installation Category; Safety Instructions - koban KMD-16 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

KMD-16
TRMS Multimeter

2. PER IEC1010 Overvoltage Installation Category

OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to
limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed
installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent
connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment

3. Safety Instructions

This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully
followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
88
www.grupotemper.com

Publicité

Table des Matières
loading

Table des Matières