General Troubleshooting
LED indicator
Normal Operation
Solid green LED
Solid yellow LED
Solid red LED
Green LED flashing
Error Codes
Orange and green
LEDs alternate
flashing
Red LED flashing
(rapidly)
Red LED flashing
(slowly)
Red LED (alternates
long and short
flashing)
Possible Cause
Greater than 90%
capacity remaining
Remaining capacity is
between 10% - 90%
Remaining capacity is
less than 10%
Battery is charging
The Co-Pilot expansion
unit was not correctly
connected to the
primary Pilot battery.
Incorrect Pilot battery
being used
Incorrect Pilot battery
being used
The environmental
temperature is too
high for the battery to
operate safely
Incorrect Pilot battery
being used
Solution
The Co-Pilot expansion unit must
be connected to a primary Pilot
battery within the first 30
seconds of being switched on
Ensure that the primary Pilot
unit is the correct model for
your Co-Pilot unit. The Co-Pilot-
12 must be connected to a Pilot-
12 and a Co-Pilot-24 must be
connected to a Pilot-24
Ensure that the primary Pilot
unit is the correct model for
your Co-Pilot unit. The Co-Pilot-
12 must be connected to a Pilot-
12 and a Co-Pilot-24 must be
connected to a Pilot-24
Ensure that the battery is placed
where there is no excess heat.
Do not rest the battery on top
of radiators or electronic
devices that may emit heat. Do
not attempt to use the battery
in temperatures exceeding 40°
C / 104° F
Ensure that the primary Pilot
unit is the correct model for
your Co-Pilot unit. The Co-Pilot-
12 must be connected to a Pilot-
12 and a Co-Pilot-24 must be
connected to a Pilot-24
7