HACH LANGE Pocket Colorimeter II Single Manuel D'utilisation page 29

Table des Matières

Publicité

All manuals and user guides at all-guides.com
5. When "S0" shows on the display, push
concentration value of the first data pair, then push
Note: Push
6. When "A0" shows on the display, push
absorbance value of the first data pair, then push
the display.
7. Do steps
5–6
8. The calibration is completed with two data pairs. If additional data
pairs are necessary for calibration:
a. When "Add" shows, push
b. Do steps
9. Push
two times to go back to measurement mode.
Remove a calibration point
To remove a calibration point from a user-entered calibration curve:
1. Set the instrument to the range to calibrate. Refer to
instrument
on page 11.
2. Push and hold
Note: If "USER" and "CAL" do not show, the factory calibration cannot be
changed on the selected range.
3. Push
until "EDIT" shows, then push
Note: Calibration points can also be removed in calibration (CAL) mode.
4. Push
until the calibration point to remove shows (i.e., S0 or S1),
then push
5. Push
until "dEL" shows, then push
Note: The minimum number of data pairs is two. When only two data pairs
remain, no more data pairs can be removed.
6. Push
two times to go back to measurement mode.
24 English
to go to the next digit.
again to enter the second data pair (S1 and A1).
5–6
again to enter more data pairs.
until "USER" and then "CAL" shows.
.
. Push
. Push
.
.
.
to enter the
.
to enter the
. "S1" shows on
Configure the

Publicité

Table des Matières
loading

Table des Matières