Correct Use; Set-Up And Commissioning - IKA R 474 Mode D'emploi

Masquer les pouces Voir aussi pour R 474:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Unpacking
- Please unpack the device carefully
- In the case of any damage a detailed report must be set im-
mediately (post, rail or forwarder
Use
The T 653 telescopic stand is used solely for mounting T 65 D
ULTRA-TURRAX
®
dispersion units.
The R 474 telescopic stand is used solely for mounting IKA
47 D agitator units.
Setting up the stand
The stand pillar is inserted into the baseplate so that the guide
groove is at the back. The black handwheel bolt is then on the
right. The stand pillar is then secured from below with a hexagon
nut, toothed lock washer and a large shaped washer.
All four rubber feet are height-adjustable, so the stand can be set
up so it does not rock, even on uneven floors.
In addition, a hole is provided in the baseplate of the R 474 tele-
scopic stand, for screwing on a 16 mm diameter support rod. A
clamp can be fixed on this rod.
All manuals and user guides at all-guides.com
Unpacking
• Delivery scope
For technical reasons associated with shipping, the telescopic
stand is supplied stripped down, as pre-assembled modules..

Correct use

The safety of the user cannot be guaranteed if the appliance is
operated with accessories that are not supplied or recommended
by the manufacturer or if the appliance is operated improperly
®
RW
contrary to the manufacturer's specifications or if the appliance is
modified by third parties.

Set-up and commissioning

Mounting the unit
The IKA
dispersion unit are placed on the arm of the stand and bolted on
from below with the bolts supplied. The help of an assistant is nee-
ded for screwing on the T 65 D ULTRA-TURRAX
The cable clips included in the delivery package are required for
fixing the power cable of the unit.
Safety disconnection device
The height-adjustable telescopic stand must be fitted with a safe-
ty disconnection device. We recommend installation of the SI 400
safety limit switch together with the SI 474 clamp support. You will
find information on their installation in the operating instructions
of the units.
Commissioning
The stand can be raised and lowered by loosening the handwheel
bolt. The integral gas shock absorber helps to keep the manual
force needed for this as light as possible. The manual force must
be applied close to the stand pillar, so that the stand can be raised
and lowered with only a slight use of force. If the force for raising
or lowering is applied at a distance from the stand pillar, a turning
moment results, leading to increased friction on the pillar guide. As
a result, the stand can only be moved with great difficulty, or can-
not be moved at all. Besides this, the danger of the stand toppling
over is increased.
When the unit is at the correct height, the handwheel bolt is tigh-
tened by hand. Check the tightness of the handwheel bolt regular-
ly when using the unit, and retighten it if necessary.
Telescopic stand
- Telescopic stand corresponding to
the type ordered T 653 oder R 474
- Operating instructions
RW 47 D agitator unit or the T 65 D ULTRA-TURRAX
®
®
dispersion unit.
®
6

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

T 653

Table des Matières