10. Measurement- Operating-Ranges and Accuracy
Temperature Measurement
TNx and THx
(Logger with ELPROsensor
typical;
Logger at room temperature)
TPx
(Logger only;
Logger at room temperature)
Relative Air Humidity Measurement
THx
Operating Ranges
Datalogger ECOLOG
rH/T Sensor; Internal
rH/T Sensor; External
For further details see: ECOLOG data sheet EZ2001E
Save Operation Area of rH/T sensor 800640: Sensor with extended temperatur range up to 100°C
Temperature (°C)
Temperature (°C)
Seite - 24
Getting Started
ECOLOG
Range
Resolution
-50°C..-25°C
0.1°C
-24.9°C..-0.0°C
0.1°C
+0.1°C..+30°C
0.1°C
+30.1°C..+70°C
0.1°C
+70.1°C..+100°C
0.1°C
+100.1°C..+140°C
0.1°C
Range
Resolution
-200°C..-100°C
0.2°C
-99.9°C..+400°C
0.1°C
+400.1°C..+500°C
0.1°C
+500.1°C..+550°C
0.2°C
Range
Resolution
0%..100%rH
0.2%rH
-35°C..55°C, display readable down to -20°C
0%..100%rH, with condensation
800637, 800639
-35°C..+55°C
0%..100%rH, with condensation
800637, 800639
-35°C..70°C
0%..100%rH, with condensation
800640
-35°C..100°C (permanent operation)
1 minute up to 110°C
0%..100%rH, with condensation
Save operation area of the rH/T sensor 800640 for extended temperature
application corresponds to the plot shown.
110°C for temperature peeks, permanent operation 100°C only.
Attention
Temperature range for the sensor cable is -35°C..80°C
The cable should not be moved at such high temperatures!
Sensor 800640 temperature dependence of the measurement error at the
time of shipment.
For sensor 800637 reduced temperature range:
-35°C..70°C only
Accuracy (k=2)
± 0.4°C
± 0.3°C
± 0.2°C
± 0.3°C
± 0.4°C
± 0.7°C
Accuracy (k=1)
± 0.6°C
± 0.3°C
± 0.3°C
± 0.5°C
Accuracy
At room temperature, 23°C: ± 1.5%rH
Hysteresis 10-90-10%rH: <1%rH
D-EZ-6003Bh
11. 2016