The most commonly used brass alloys used for bodies and other brass
components contain 2-3% of lead (Pb), EC no. 231-468-6, CAS no. 7439-
92-1. The lead will not be released to the gas or surrounding environment
during normal use. After end of life the product shall be scrapped by an
authorized metal recycler to ensure e cient material handling with minimal
impact to environment and health.
To date we have no information that indicates that other materials
containing SVHC of concentrations exceeding 0.1% are included in any
GCE product.
9.2. MAINTENANCE
9.2.1. MAINTENANCE
GCE recommends the product owner to make regular visual inspections
at least once a year, including checks for tightness and right functioning
of the terminal unit.
Maximum usability time within the framework of the maintenance shall be
10 years.
a) Base valve kit for all gasses
(to be replaced every 5 years)
Ref.: K292404
b) Valve-fi lter unit
(to be replaced every year)
Ø 6 (mix N2O+O2; O2+N2; AIR 800) Ref.: SPK36810087 (1 pcs)
Ø 6 (mix N2O+O2; O2+N2; AIR 800) Ref.: SPK36810088 (50 pcs)
Ø 7 (O2; Air; N2O; CO2) Ref.: SPK36810038 (1 pcs)
Ø 7 (O2; Air; N2O; CO2) Ref.: SPK36810049 (50 pcs)
Ø 8 (Vac; O2+CO2; N2) Ref.: SPK36810040 (1 pcs)
Ø 8 (Vac; O2+CO2; N2) Ref.: SPK36810050 (50 pcs)
The design of the medical socket permits very simple and rapid
maintenance without disassembly of the valve-fi lter unit.
Tools required for mounting and maintenance of all the sockets :
Multifunction key K 00 70 91.